Kaye ValProbe Standard Loggers
Easy-to-Use Wireless Thermal Validation Solution for Standard Processes
Description
The ValProbe® system's wireless probe design eliminates the need for hard-wired sensors, simplifying access to hostile, remote or hard to reach environments. The ValProbe system is ideally suited for applications where high measurement accuracy and regulatory compliance are priorities:
- Pharmaceutical Processing
- Medical Device Sterilization
- Food Processing
- Environmental Monitoring
Built in data processing and reporting capabilities extend the ValProbe systems operating convenience far beyond mere data acquisition. The ValProbe system performs calculations and generates custom user-defined reports for up to 200 sensors at one time. Graph reports can include many sensors and limits for easy review of study data. User calibration and verification is a quick and easy process with the new CTR-40 Temperature Reference.
ValProbe satisfies FDA Regulation 21 CFR Part 11 requirements for electronic signatures and records and complies with ISO-17665, EN285 and HTM-2010 for saturated steam sterilization.
For extreme low temperature check out the Kaye ValProbe® Cryo Logger with a Temperature range from -85°C to 140°C
Features
- RTD technology delivers unrivaled measurement accuracy over a wide operating range
- System downloads up to 10 probes simultaneously
- Easily defined cycle-based data collection, calculation, and reporting from up to 200 sensors
- Enables compliance with FDA Regulation 21 CFR Part 11
- Operates from -45C to 140°C, and up to 10 BAR absolute
- Designed for easy on-site calibration
- Battery life indicator and field replaceable battery
- Available as a economical validation starter kit complete with software, single reader, two temperature and one pressure logger (P/N V2543).
- Reporting capability enhanced with up to 25 user programmable groups.
Applications
The ValProbe system is ideally suited for applications where high measurement accuracy and regulatory compliance are priorities:
- Sterilizers
- Ovens
- Lyophilizers
- Stability Chambers
- Temperature Chambers
- Fridges, Freezers
- Warehouses
Specifications
Kaye ValProbe Generic Specifications
Reader 2 Dimensions |
320mm x 155mm x 60mm (12,60” x 6,10” x 2,36”) |
Logger Dimensions |
Height: 48mm / Diameter: 36mm (Height: 1,9’’ / Diameter: 1,4’’) |
Logger Material |
Stainless Steel 316L and Peek |
Battery |
Field replaceable – 3.6V Lithium |
Sampling Rate |
1 sec to 12 hours |
Data Storage |
10,000 Samples retained in non-volatile memory |
Real-Time Clock Accuracy |
< 15sec / day |
Calibration |
NVLAP/DAkkS Calibration |
Verification |
Automated User Verification capability |
Sensing Element |
Precision Platinum RTD |
Environmental Temperature |
-85°C/-45°C to 140°C |
Environmental Pressure |
0 – 10 bar absolute |
Environmental Humidity |
0 – 100% condensing |
Regulatory Compliance |
UL and CE |
Kaye ValProbe Temperature Loggers Specifications
Specifications |
Rigid |
Flexible |
Bendable |
Surface |
Sensor Type |
Single Sensor |
Single Sensor |
Single and Dual SENSOR |
Sensor |
Sensor Length |
1.5, 3, 6, 9” inches (38, 76, 152, 229mm) |
40” inches (1000mm) |
12, 24, 36” inches (305, 610, 915mm) |
- |
Tip Diameter |
0.118” (3mm) |
0.095” (2.4mm) |
0.095” (2.4mm) |
1.26” (32mm) |
Measurement Range |
-85°C / -45°C to 140°C |
-85°C / -45°C to 140°C |
-85°C / -45°C to 400°C |
-85°C to 140°C |
Accuracy |
-85°C to 0°C, ±0.25°C / Cryo VP; -45°C to 0°C, ±0.25°C / Standard VP; 0°C to 140°C, ±0.1°C |
-85°C to 0°C, ±0.25°C / Cryo VP; -45°C to 0°C, ±0.25°C / Standard VP; 0°C to 140°C, ±0.1°C |
±0.25°C from -85°C to 0°C / Cryo VP; ±0.25°C from -45°C to 0°C / Standard VP; ±0.1°C from 0°C to 140°C; ±0.2°C from 140°C to 250°C; ±0.5°C from 250°C to 360°C |
-85°C to 0°C, ±0.25°C; 0°C to 140°C, ±0.1°C |
Kaye ValProbe Pressure, Temperature and Humidity Logger Specifications
Specifications |
Pressure & Temp |
Humidity & Temp |
Environmental Temperature |
0°C to 140°C |
0°C to 90°C |
Sensor Type |
Single P/T Sensor |
RH: EMD4000 |
Tip Diameter |
1/4” NPT |
Diameter 18mm (0,71”) Length 35mm (1,38”) |
Measurement Range |
0°C to 140°C / 0 – 5bar abs. |
25% to 85% RH |
Accuracy |
0°C to 120°C; ±25mb 120°C to 135°C; ±10mb 135°C to 140°C; ±25mb 0°C to 140°C; ±0.1°C |
25% to 85% RH; ± 2% RH (at 25°C to 40°C) 0°C to 90°C; ±0.1°C |
Documents