KAYE IRTD 400 - Temperature Standard Probe - Temperature Reference Standard
High Accurate Traceable Temperature Reference Standard
Description
The IRTD is a high accuracy traceable Temperature Standard probe used as part of the Kaye Validation systems for Calibrating Temperature Sensors. During calibration/ verification, sensors are compared against the IRTD and offsets are automatically calculated and stored. The broad temperature ranges from Ultra Low Temperature (-196°C) to High Temperature (+420°C), enables it to be used in virtually all temperature validation applications.
Temperature Reference Standard
The IRTD is an integral part of Kaye's automated sensor calibration system and communicates directly with all Kaye Validation systems. With an Accurate Temperature accuracy of +/- 0.025, the IRTD temperature Standard Probe provides a 4 to 1 ratio for calibrating sensors to 0.100°C. With this accuracy this Temperature Reference Standard is one of the most accurate industrial Temperature Reference Standards in the market.
The Temperature Reference Standard probe can be used for confirming the accuracy of other temperature devices (Thermometer Calibration, calibration of other RTDs or Thermocouples). For this purpose, Kaye has developed IRTDWin® software. The IRTDWin® software is a convenient interface screen that lets you track probe stability, numerically and graphically, and confirm system accuracy.
See real-time data of your IRTD using the calibration interface and the Kaye IRTD Console software. The Calibration Standard interface connects directly to your computer using USB and to up to two IRTDs. The software immediately shows the temperature of the IRTDs with an accurate resolution to the thousandth degree. The interface is simple, easy to use, and does not need a power source, as it is powered by the computer.
IRTD Display
The Kaye IRTD Display is a standalone interface with a touchscreen. This advanced display unit seamlessly connects with up to two Kaye IRTD Reference Standard probes, offering real-time temperature data in a user-friendly format. Equipped with a 4.3" touch screen, the Kaye IRTD Display provides clear and intuitive readings for both IRTDs, allowing users to monitor temperature with ease. The display also showcases the temperature difference (Delta) between the two connected IRTDs, enabling quick and accurate analysis. Supporting both types of IRTD Connector cables, this versatile unit ensures compatibility with your preferred setup.
Features
- Temperature range -196 to 420°C
- Accuracy over range ±0.025°C
- Resolution 0.001°C
- Sensor element 200 Ohm Platinum sensor
- Sheath material Inconel™ 600
- Immersion depth 101.6mm (4")
- Calibration Traceable to NIST or PTB; Recommended calibration period is 1 year
- Measurement rate 30 readings per second
- Ambient temperature range 0 to 60°C (32 to 140°F)
- Humidity 0 to 95% non-condensing
- Dimensions Overall length: 603mm (23.75"); Grip: 89mm x 32mm (3.5" x 1.25") Sensor sheath: 457mm x 6.35mm (18" x 0.25")
Applications
This temperature standard probe is used for applications ranging from -196°C to 420°C and where traceability and high accuracy is required
Specifications
Specifications | IRTD-400 |
Temperature Range | -196 to 420°C |
Accuracy Over Range* | 0.025°C |
Resolution | 0.001°C |
Sensor Element | 200 Ohm Platinium RTD Sensor |
Sheath Material | Inconel 600 |
Minimum Immersion Depth | 101.6 mm (4") |
Calibration ** | Traceable to NIST, DAkkS calibrated (Europe); recommended calibration period is 1 year |
Power to Probe | Unregulated DC, 10 to 25V, first probe: 850mW at 15V; each additional probe 550mW |
Power Supply *** | Adapter: 110 VAC US-style, or 230 VAC VDE-approved |
Measurement Rate | 1 reading per second |
Environmental | Ambient Temperature Range: 0 to 60°C (32 to 140°C) Humidity 0 to 95% non condensing |
Dimensions | Overall length 603mm (23.75") Grip: 89mm x 32mm (3.5" x 1.25") Sensor sheath: 457mm x 6.35mm (18" to 0.25") |
* Accurate for 1 year, 0 to 60°C ambient. Includes calibration certificate with traceable to NIST
** Kaye provides a recertification service for calibrating the Temperature Standard
*** Power Supply is not required for use with the ValProbe system