Kaye ValProbe RT Humidity/Temperature Logger
Humidity Measurement from 15 % to 95 % and Temperature Range from 0°C to 70°C
Description
The ValProbe RT logger with humidity and temperature sensors comes with RF technology to give users real-time data viewable on their Validation Console. With a humidity range of 15% to 95%, and a temperature range from 0°C to 70°C, this logger has a wide range of capabilities for temperature and humidity mappings.
Digital sensor technology delivers unrivalled measurement accuracy and the new logger electronic design guarantees long battery life. Operator programmable sample rates starting at 1 second, allow collection of up to 100,000 data points per sensor.
FIELD-REPLACEABLE DIGITAL SENSOR
We designed this humidity logger with the end user in mind. At Kaye we understand that sending loggers back to the manufacturer for factory calibration, verifications, or repairs can slow down your production. That’s why this new ValProbe RT logger is built with a field-replaceable digital sensor. This allows users to quickly and cost-effectively replace the sensor without having to send the entire logger back to the calibration lab. The sensors are pre-calibrated before being sent to the user, guaranteeing precision and accuracy right from the start.
EASY VERIFICATIONS IN YOUR LAB
When verifying for temperature, simply attach the 1-meter extendable cable sensor for easy verification in a drywell, like the Kaye LTR-150. This dry block allows users to verify up to 18 humidity/temperature ValProbe RT loggers at once, saving a substantial amount of time. Once the probe is inserted into the dry block, the ValProbe RT software will perform an automatic verification and verify the temperature at multiple set points, determining the accuracy of the logger.
Verifying relative humidity is also easy by using a portable humidity test chamber, which will evaluate the performance of multiple ValProbes at a time. The RT software will perform an automatic verification of the humidity sensors to show if the loggers have drifted. Additionally, reports can be generated automatically inside the ValProbe RT software, simplifying the process to review the report of the loggers.
Features
- High accuracy measurement: 2% on humidity and 0.15°C for temperature
- 100,000 data samples per sensor
- Field-replaceable digital temp/humidity sensor
- 1 meter sensor extension cable
- Economical field-replaceable battery
- Operator programmable sample rate, start, delay and stop function
Applications
- Stability chambers
- Incubators
- Warehouses
- Temperature chambers
Specifications
Specifications |
Humidity/Temperature |
Sensor Type |
Digital RH/T Sensor |
Sensor Length |
Up to 1 meter |
Tip Diameter |
Sensor tip: 0.295”, 7mm Outer filter: 0.472”, 12mm |
Measurement Range |
Temp: 0°C to 70°C Humidity: 15% to 95% |
Accuracy |
0°C to 70°C, ±0.15°C 15% to 95% RH, 2% |
Kaye ValProbe RT Generic Specifications |
|
Logger Dimensions |
Hight: 1,9'' / Diameter 1,4'' |
Logger Material |
Stainless Steel 316L and Peek |
Battery |
Field replaceable - 3.6V Lithium |
Sampling Rate |
Starting at 1 sec |
Data Storage |
100.000 Samples retained in non-volatile memory |
Real-Time Clock Accuracy |
< 15sec/day |
Calibration |
NVLAP/DAkkS Calibration |
Verification |
Automated User Verification capability |
Sensing Element |
Digital Sensor |
Environmental Temperature |
0°C to 70°C |
Environmental Humidity |
Non-condensing |
Regulatory Compliance |
UL and CE |