Kaye ValProbe Freeze Dryer Logger
Special Design to achieve Highest Accuracy for Surface Temperature Measurement
Description
The X2534 ValProbe surface temperature logger design provides enhanced and precision wireless measurement capability ideal for freeze dryer shelf temperature measurement over an extended temperature range from -85°C to +140°C.
The embedded RTD technology is integrated into a unique compact surface design, delivering unrivaled measurement accuracy, stability, and response time over the full temperature range. In addition, the Freeze Dryer Logger has been designed to minimize power consumption to deliver maximum battery performance in extremely cold environments.
The new surface Logger is compatible with the existing ValProbe family of Loggers, Readers and software. It is designed for easy on-site user verification between studies with the traditional IRTD and the low temperature CTR-80 bath. Using the standard Kaye Report Wizard, Validation engineers are guaranteed consistent reports among all Kaye products with the capability to merge study data with a Kaye Validator or the RF ValProbe. The surface Logger and Report Wizard offer added value when performing validation execution, saving time and money.
The ValProbe system satisfies FDA Regulation 21 CFR Part 11 requirements for electronic signatures, and records and complies with DIN ISO 17665 for saturated steam sterilization and is used to fulfill GMP requirements.
Features
Features of the X2534
- Unique integrated design with embedded RTD technology
- Wireless sensing design that maximizes thermal response time and performance
- Self-adhering thermal pad insures continuous thermal conductivity during vacuum cycles for ultimate surface measurement performance
- Temperature range for complete logger: -85°C to 140°C
- Battery life performance — Designed for maximum battery life at extremely cold conditions
- Fully compatible with the full family of ValProbe Loggers, Readers and Software
Applications
- Freeze Dryer
- Cryogenic Vessels
- Freezers
- Incubators
- Warehouses
- Sterilizers
- Surface Measurement
Specifications
Data Logger Specifications
Sensing Element | Precision Platinum RTD |
Measurement Range and Accuracy | 0° to 140°C ±0.1°C -85° to 0°C ±0.25°C |
Environmental Temperature Humidity Pressure |
-85° to 140°C 0% to 100% humidity, condensing 0 to 10 bar absolute (0 to 130 psia) |
Logger Material |
316L stainless steel; |
Logger Dimensions | Height 2.28 in x 1.41 in diameter (58 mm x 36 mm) |
Sensing Area Dimension | 1.34 in (34 mm) diameter |
Battery | Field-replaceable 3.6 V lithium thionyl chloride 1600 hours at 1 sec at 20°C/180 hours at 1 sec at -85°C |
Sampling Rate | 1 second to 12 hours |
Data Storage | 10,000 samples retained in non-volatile EEPROM memory |
Calibration | Factory calibrated (NVLAP accredited) with user calibration capability |
Real Time Clock Accuracy | 15 seconds per 24 hours (0.0174%) |
Regulatory Compliance | UL and CE |
Thermal Contact Pad | Die cut thermal pad with natural inherent tack to logger |