Kaye ValProbe Standard Loggers
Easy-to-use wireless thermal validation solution for standard processes
{tab Description}
The wireless probe design of the ValProbe® system eliminates the need for wired sensors, simplifying access to harsh, remote or hard-to-reach environments. The ValProbe system is ideal for applications where high measurement accuracy and regulatory compliance are priorities:
- Pharmaceutical processing
- Sterilization of medical devices
- Food processing
- Environmental monitoring
Integrated data processing and reporting capabilities extend the operating convenience of ValProbe systems beyond mere data acquisition. The ValProbe system performs calculations and generates user-defined custom reports for up to 200 sensors at a time. Graph reports can include many sensors and limits for easy review of study data. User calibration and verification is a quick and easy process with the new CTR-40 temperature reference.
ValProbe meets the requirements of FDA 21 CFR Part 11 for electronic signatures and registers and complies with ISO-17665, EN285 and HTM-2010 for saturated steam sterilization.
For extremely low temperatures, see the Kaye ValProbe® Cryo Logger with a temperature range of -85 ° C to 140 ° C
{tabCaracterísticas}
- La tecnología RTD ofrece una precisión de medición sin igual en un amplio rango operativo
- El sistema descarga hasta 10 sondas simultáneamente.
- Fácilmente definido recopilación de datos basada en el ciclo, cálculo e informes de hasta 200 sensores
- Permite el cumplimiento de la normativa FDA 21 CFR Parte 11
- Funciona desde -45 ° C hasta 140 ° C, y hasta 10 BAR absolutos
- Diseñado para una fácil calibración en el sitio
- Indicador de duración de la batería y batería reemplazable en el campo
- Disponible como un kit de inicio de validación económica completo con software, lector único, dos registradores de temperatura y uno de presión (P / N V2543).
- Capacidad de informes mejorada con hasta 25 grupos programables por el usuario.
{tab Aplicaciones}
El sistema ValProbe es ideal para aplicaciones donde la alta precisión de medición y el cumplimiento normativo son prioridades:
- Esterilizadores
- Hornos
- Liofilizadores
- Cámaras de estabilidad
- Cámaras de temperatura
- Frigorificos, Congeladores
- Almacenes
{tab Especificaciones}
Kaye ValProbe Generic Specifications
Reader 2 Dimensions |
320mm x 155mm x 60mm (12,60” x 6,10” x 2,36”) |
Logger Dimensions |
Height: 48mm / Diameter: 36mm (Height: 1,9’’ / Diameter: 1,4’’) |
Logger Material |
Stainless Steel 316L and Peek |
Battery |
Field replaceable – 3.6V Lithium |
Sampling Rate |
1 sec to 12 hours |
Data Storage |
10,000 Samples retained in non-volatile memory |
Real-Time Clock Accuracy |
< 15sec / day |
Calibration |
NVLAP/DAkkS Calibration |
Verification |
Automated User Verification capability |
Sensing Element |
Precision Platinum RTD |
Environmental Temperature |
-85°C/-45°C to 140°C |
Environmental Pressure |
0 – 10 bar absolute |
Environmental Humidity |
0 – 100% condensing |
Regulatory Compliance |
UL and CE |
Kaye ValProbe Temperature Loggers Specifications
Specifications |
Rigid |
Flexible |
Bendable |
Surface |
Sensor Type |
Single Sensor |
Single Sensor |
Single and Dual SENSOR |
Sensor |
Sensor Length |
1.5, 3, 6, 9” inches (38, 76, 152, 229mm) |
40” inches (1000mm) |
12, 24, 36” inches (305, 610, 915mm) |
- |
Tip Diameter |
0.118” (3mm) |
0.095” (2.4mm) |
0.095” (2.4mm) |
1.26” (32mm) |
Measurement Range |
-85°C / -45°C to 140°C |
-85°C / -45°C to 140°C |
-85°C / -45°C to 400°C |
-85°C to 140°C |
Accuracy |
-85°C to 0°C, ±0.25°C / Cryo VP; -45°C to 0°C, ±0.25°C / Standard VP; 0°C to 140°C, ±0.1°C |
-85°C to 0°C, ±0.25°C / Cryo VP; -45°C to 0°C, ±0.25°C / Standard VP; 0°C to 140°C, ±0.1°C |
±0.25°C from -85°C to 0°C / Cryo VP; ±0.25°C from -45°C to 0°C / Standard VP; ±0.1°C from 0°C to 140°C; ±0.2°C from 140°C to 250°C; ±0.5°C from 250°C to 360°C |
-85°C to 0°C, ±0.25°C; 0°C to 140°C, ±0.1°C |
Kaye ValProbe Pressure, Temperature and Humidity Logger Specifications
Specifications |
Pressure & Temp |
Humidity & Temp |
Environmental Temperature |
0°C to 140°C |
0°C to 90°C |
Sensor Type |
Single P/T Sensor |
RH: EMD4000 |
Tip Diameter |
1/4” NPT |
Diameter 18mm (0,71”) Length 35mm (1,38”) |
Measurement Range |
0°C to 140°C / 0 – 5bar abs. |
25% to 85% RH |
Accuracy |
0°C to 120°C; ±25mb 120°C to 135°C; ±10mb 135°C to 140°C; ±25mb 0°C to 140°C; ±0.1°C |
25% to 85% RH; ± 2% RH (at 25°C to 40°C) 0°C to 90°C; ±0.1°C |
{tab Documentos}
Inglés
{tab Video}
{/ tabs}