Kaye ValProbe Pressure/ Temperature Logger
Pressure measurement for pharmaceutical and medical device applications
Description
The combined pressure/temperature ValProbe data logger provides an integrated solution to small scale sterilizer validation, monitoring and routine control. The data logger is ideal for benchtop or lab sterilizers, hospital validation and routine control, and other sterilization validations needs in dental offices and other medical device industries.
Features
- Combined pressure and temperature sensor
- High accuracy temperature and pressure measurement to meet regulatory guidelines
- 10,000 data sample memory
- Cost effective, field-replaceable battery
- Operator programmable sample rates from 1 sec to 12 hours
- Configurable start, change, and stop events
- Enables compliance with ISO-17665, EN 554, EN 285 and HTM-2010
Applications
- Steam sterilizer validation
- Hospital validation and routine control
- Small-scale sterilization, like benchtop sterilizers in dental offices
- Parametric release applications
Specifications
Sensing Element Temperature |
Pressure Platinum RTD |
Measurement Range and Accuracy |
0°C to 140°C ±0.1°C |
Environmental Temperature Environmental Humidity Environmental Pressure |
-45°C to 140°C 0 to 100% humidity, condensing 0 to 10 bar absolute (0 to 147 psia) |
Logger Material |
316 stainless steel |
Battery |
Field-replaceable 3.6V Lithium |
Sampling Rate |
1 second to 12 hours |
Data Storage |
5000 samples per sensor retained in EE PROM memory |
Calibration |
Factory calibrated (NIST-traceable) with user calibration |
Real time clock accuracy |
15 seconds per 24 hours (0.0174%) from -45 to 140°C |
Regulatory Compliance |
UL, CE, Intrinsically Safe |
Sensing Element Pressure |
Absolute Pressure Sensor |
Operating Range |
0 to 5 bar absolute (0 to 73 psia) |
Accuracy |
Temperature Range, Accuracy >0°C to 120°C, ±25 mb accuracy >120°C to 135°C, ±10mb accuracy >135°C to 140°C, ±25mb accuracy |
Logger Dimensions |
Diameter 46mm x 35 mm (1 7/8 in x 1 3/8 in) Height 72 mm (2 7/8 in) |
Calibration |
Factory calibrated (NIST-traceable / DAkkS tracable) with user calibration capability |
Part Number |
x2532 |
Documents